corresponding settings according to the following procedure:
"SYS PROMPT" and "DIA PROMPT" can control closing or unsealing of the SYS and DIA prompt
separately.
The prompt is on or off according to the high and low limits which have been set up. When the
measure result is higher than the high limit or lower than the low limit, and meanwhile, the
"PROMPT SOUND" is on, "SYS PROMPT" or "DIA PROMPT" accordingly on, the prompt will
occur.
The adjustable ranges of the high and low limits of the adult mode prompt are as follows:
SYS PROMPT: 40~270 mmHg
DIA PROMPT: 10~215 mmHg
The adjustable ranges of the high and low limits of the pediatric mode prompt are as follows:
SYS PROMPT: 40~200 mmHg
DIA PROMPT: 10~150 mmHg
The adjustable ranges of the high and low limits of the neonatal mode prompt are as follows:
SYS PROMPT: 40~135 mmHg
DIA PROMPT: 10~100 mmHg
"DEFAULT" includes the main content:
Measure mode: adult;