80
range
Inflation
adult
160mmHg(21.33 kPa)
pediatric 120mmHg(16kPa)
neonatal
70mmHg(9.33 kPa)
Prompt Range
adult mode
SYS PROMPT: 40~270 mmHg(5.3~36.0 kPa)
DIA PROMPT: 10~215 mmHg(1.3~28.7 kPa)
pediatric mode
SYS PROMPT: 40~200 mmHg(5.3~26.7 kPa)
DIA PROMPT: 10~150 mmHg(1.3~20.0 kPa)
neonatal mode
SYS PROMPT: 40~135 mmHg(5.3~18.0 kPa)
DIA PROMPT: 10~100 mmHg(1.3~13.3 kPa)
Overpressure protect
adult mode
297
±
3 mmHg
39.6
±
0.4 kPa
pediatric mode
240±5 mmHg
32
±
0.67kPa
neonatal mode
147
±
3 mmHg
19.6
±
0.4 kPa
Resolution
Pressure
1 mmHg
0.133 kPa
Pulse Rate
±5 %
Measurement Accuracy
Cuff Pressure
Static pressure: ±3 mmHg(±0.4 kPa)