81
Accuracy
Error
The BP value measured by the device is equivalent with the
measurement value of Stethoscopy, perform clinical verification in
accordance with the requirements in ISO 81060-2: 2013, whose
error meets the followings:
Maximum mean error: ±5 mmHg
Maximum Standard deviation: 8 mmHg
Operating
temperature/
humidity
+5 ºC~40 ºC 15 %RH~85 %RH(Non-condensing)
Transport
Transport by general vehicle or according to the order contract,
avoid pounded, shake and splash by rain and snow in
transportation.
Storage
Temperature: -20 ºC~+55 ºC; Relative humidity: ≤95 %; No
corrosive gas and drafty.
Atmospheric pressure
700 hPa~1060 hPa
Power Supply
DC 3 V
Battery service life
When the temperature is 23 ºC, limb circumference is 270 mm, the
measured blood pressure is normal, 2 "AA" alkaline batteries can