●In order to avoid the accumulation of electrostatic charge, it is recommended to store, maintain and
use the equipment at a relative humidity of 30% or more. The floor should be covered with ESD
dissipated carpets or similar materials. In the use of the components, non-synthetic clothing should
be wore.
●In order to prevent electrostatic discharging to the ESD-sensitive parts of the device, the personnel
should contact the metal frame of the components or the large metal objects near the device. When
using the device, especially when it is possible to contact the ESD-sensitive parts of the device, the
operator should wear a grounded bracelet designed for ESD-sensitive devices. For more information
on proper use, please refer to the instructions provided with the bracelet.
●All potential users are advised to understand the ESD warning symbols and receive training on
ESD precautions.
●The most basic content of the ESD precautionary procedure training should include an introduction
to electrostatic charge physics, voltage level in the conventional case, and damage to the electronic
components when the operator with electrostatic charge contacts them. In addition, the methods for
preventing electrostatic buildup, and the manner and reasons for the release of human body static
electricity to the ground or equipment frame or the use of a bracelet to connect the human body to
the equipment or the ground before establishing the connection should be described.