EasyManuals Logo

Gossen MetraWatt PROFITEST MBASE MTECH User Manual

Gossen MetraWatt PROFITEST MBASE MTECH
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #31 background imageLoading...
Page #31 background image
GMC-I Messtechnik GmbH 31
10.4 Selective Earth Resistance Measurement with Accessory Current Clamp Sensor
As an alternative to the classical measuring method, it is possible to perform measurements with a current clamp sensor.
Selective Earth Resistance Measurement (Mains Powered) – Connection Circuit Diagram
Key figure below
R
B
Operational earth
R
E
Earth resistance
R
L
Conductor resistance
R
X
Earth resistance via equipotential bonding systems
R
S
Probe resistance
PAS Equipotential bonding strip
RE Overall earth resistance (R
E1
// R
E2
// water pipe)
Measurement without clamp: R
E
= R
E1
// R
E2
Measurement with clamp: R
E
= R
E2
=
Select the Measuring Function
Connection
Connected components: 2-pole adapter, clamp sensor and probe
Set parameters at the test instrument
q Connection type: 2-pole adapter + clamp sensor
After parameter selection: automatic setting to 10
measuring range and transformer ratio of 100 mV/A
q Transformer ratio of current clamp sensor: see table below
q Measuring range (test current selection):
10 k (4 mA), 1 k (40 mA), 100 (0.4 A), 10 (> 0.65 A)
The DC + function can be selected for systems with
RCCB.
q Contact voltage: UL < 25 V, < 50 V, < 65 V, freely selectable
voltage, see section 5.7
Setting Parameters at the Current Clamp Sensor
q Measuring range of current clamp sensor: see table below
Select the Measuring Range at the Current Clamp Sensor
Important Instructions Regarding the Use of the Current Clamp Sensor
Use only the METRAFLEX P300 or Z312A current clamp
sensor for this measurement.
Read and adhere strictly to the operating instructions for the
METRAFLEX P300 current clamp sensor, as well as the safety
precautions contained therein.
Be sure to observe the direction of current flow (see arrow on the
current clamp sensor).
The clamp must be securely attached for use. The sensor must
not be permitted to move during measurement.
The current clamp sensor may only be used at an adequate
distance from strong, extraneous fields.
Before use, always inspect the electronics housing, the
connector cable and the flexible current sensor for damage.
In order to prevent electrical shock, keep the surface of the
METRAFLEX clean and free of contamination.
Before use, make sure that the flexible current sensor, the
connector cable and the electronics housing are dry.
P
R
O
F
I
T
E
S
T
w
a
t
e
r
p
i
p
e
SE
2
E
1
B
U
Sonde
I
Zange
-----------------


R
E
Tester METRAFLEX P300 clamp Tester
Transforma-
tion Ratio
Parameter
Switch Measuring
Range
Measuring
Range
1:1
1 V / A
3 A (1 V/A) 3 A
0.5 ... 100
mA
1:10
100 mV / A
30 A (100 mV/A) 30 A 5 ... 999 mA
1:100
10 mV / A
300 A (10 mV/A) 300 A 0.05 ... 10 A

Table of Contents

Other manuals for Gossen MetraWatt PROFITEST MBASE MTECH

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Gossen MetraWatt PROFITEST MBASE MTECH and is the answer not in the manual?

Gossen MetraWatt PROFITEST MBASE MTECH Specifications

General IconGeneral
BrandGossen MetraWatt
ModelPROFITEST MBASE MTECH
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals