3-16
Settings and Measurement
Auto Save ........................... When this parameter has been selected, a file is created for each data
capture time.
Off, 1, 3, 6, 12, 24 hours
The maximum data capture time that can be selected depends on the
sampling interval.
• 24 hours: 1s or less sampling interval
• 12 hours: 500 ms or less sampling interval
• 6 hours: 200 ms or less sampling interval
• 3 hours: 100 ms or less sampling interval
If the allowable data capture time is exceeded, the actual data capture
time will be shortened.
Statistical Calculation
Statistical calculation is performed on the replay data between the cursors.
Function .............................. There are five types of between-cursor calculation functions, and four
of these can be selected.
Off: Calculation is not performed.
Average: The simple average value of the data between the cursors is
displayed.
Max: The maximum value of the data between the cursors is
displayed.
Min: The minimum value of the data between the cursors is
displayed.
Peak: The peak value of the data between the cursors is displayed.
RMS: The effective value of the data between the cursors is
displayed.
R.M.S = ΣD
2
/n
D: data n: number of data
Channel specification: Specifies the two channels "A" and "B" for
calculation.
CHECKPOINT
Calculation results during data capture are displayed on the Calculation
Display screen.