Settings and Measurement
3-17
rr
rr
r ALM Settings Window
The Alarm and Trigger settings are made here.
Combination
Determines the alarm output conditions. The logic alarm cable (B-513, option) is required for alarm
output.
OR: Outputs an alarm when each alarm condition is met.
AND:Outputs an alarm when all of the alarm conditions are met.
Repeat Trigger
Used to specify repeated measurement. Measurement is performed repeatedly according to the
Trigger Start and Stop conditions until the START/STOP key is pressed.
On: Repeated measurement
Off: Single measurement operation
CHECKPOINT
If repeated measurement has been specified and Memory selected as the capture destination, data
captured to the GL450's internal memory is overwritten.
Setting Selections available Setting method
Combination OR, AND ENTERSelectENTER
Repeat Trigger On, Off ENTERSelectENTER
Mode H, L, Win In, Win Out ENTERSelectENTER
Lower - Level - Upper Numeric settings ENTERSelectENTER
Output 1 to 4 ENTERSelectENTER
Pulse
• Mode L, H, Win In, Win Out ENTERSelectENTER
Lower - Level - Upper
Numeric settings ENTERSelectENTER
• Output 1 to 4 ENTERSelectENTER
Logic
• Mode X, L, H ENTERSelectENTER
• Output 1 to 4 ENTERSelectENTER
Trigger
• Start Off, Level, Alarm, Ext., Specified Time ENTERSelectENTER
Alarm output selection: 1 to 4 ENTERSelectENTER
• Stop Off, Level, Alarm, Ext., Time, Specified Time ENTERSelectENTER
Alarm output selection: 1 to 4 ENTERSelectENTER
Time setting: 1 s to 9999 h ENTERSelectENTER