3-20
Settings and Measurement
Stop .................................... Specifies the trigger condition that must be met in order to stop
measurement.
Off: No trigger is used. Instead, measurement is stopped by
pressing the START/STOP key.
Level: Measurement is stopped when the trigger condition is satisfied.
Note:The Level condition is the level value set for the specified
channel range set (voltage, temperature).
Alarm: Measurement is stopped when the alarm condition is satisfied.
Ext.: Measurement is stopped when a signal is input from an
external trigger terminal.
Time: Sets the amount of time from the start until the end of measurement.
The time can be specified from 1 second to 9999 hours.
Specified Time:
Measurement is stopped (data capture ends) at the specified time or
when the time trigger condition is satisfied.
If Repeat Trigger has been specified, the trigger action is as follows:
Triggered at: Off Absolute Time/On Relative Time