EasyManua.ls Logo

GRAPHTEC midi LOGGER GL240 - Page 76

GRAPHTEC midi LOGGER GL240
152 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-34
CHAPTER 3 Settings and Measurement
4ch Voltage / Temperature Terminal of the wireless sensor setting: optional
(GL100-WL+GS-4VT)
Perform the settings for the 4ch voltage / temperature terminal input of wireless sensors.
When the wireless sensor and 4ch voltage / temperature terminal are connected, the setting screen is
displayed.
(1)-22 Input setting
This is used to select the input conditions.
Setting Description
1ch to 4ch Off The input signal is not measured. The waveform and digital display also are not performed.
DC Used to measure the direct-current voltage.
TEMP Used to measure the temperature.
5ch to 8ch No AMP Because the other wireless sensor cannot be installed, the “No AMP” is displayed..
(1)-23 Sensor setting
This is used to select the sensor type to be used.
Setting Description
TC-K Set when the thermocouple to be connected is K-type.
TC-T Set when the thermocouple to be connected is T-type
(1)-24 Range setting
Setting Description
DC 20, 50, 100, 200, 500mV; 1, 2, 5, 10, 20, 50V; 1-5V
TEMP 200C xed range
Do not change the range through the GL100-WL after setting the range of the GL240.
The displayed values of the GL240 may not match the measured values.
(1)-25 EU setting
This is used to unit-convert the measurement signal.
Refer to “EU Function, Meas. Value, EU Output Value, and Reads the current temperature measurement
value” in “(1)-5 EU (Scaling settings)” in “3.4 Setting Menus”. In the case of Voltage, refer to “<Setting
example: For voltage input>”.

Table of Contents

Other manuals for GRAPHTEC midi LOGGER GL240

Related product manuals