7.11.2 Measuring range
1. "main menu > commissioning > measurement > on >
measuring range"
The parameters in the display depend on the selected measuring
cell.
2. Select and set all measuring ranges in succession.
ClO
2
measuring range
1. "main menu > commissioning > measurement > on >
measuring range > ClO2".
Predefined ranges
others
Enter the end of the measuring range.
Temperature measuring range
1. "main menu > commissioning > measurement > on >
measuring range > temperature"
2. Select the temperature unit.
3. Select the temperature measuring range.
pH measuring range
1. "main menu > commissioning > measurement > on >
measuring range > pH"
Predefined ranges
others
Enter the start and the end of the measuring range.
ORP measuring range
Make the settings as described before.
7.11.3 Current output
1. "main menu > commissioning > measurement > on > current
output".
Predefined ranges
2. Select the range.
others
Set the start and the end of the measuring range.
1. Enter current value for "start meas range".
2. Enter current value for "end meas. range".
The specified range for the current output is assigned to the
corresponding ClO
2
concentration.
7.11.4 ClO
2
alarm value
"main menu > commissioning > measurement > on > ClO2 alarm
> on"
1. Select the submenu.
2. Enter the value.
7.12 Calibration
Calibration can only be done, if the sample water contains ClO
2
.
For details on calibration of the measuring cell, see section
8.3 Calibration.
7.13 Water meter
"main menu > commissioning > water meter > on"
Water meter with pulse signal
1. Enter volume per pulse.
2. Enter the maximum flow.
Water meter with current signal
1. "main menu > commissioning > water meter > on > current
signal"
Predefined ranges
measuring range
ClO2
temperature
pH/ORP
ClO2
0.0... 0.50 mg/l
0.0... 1.00 mg/l
0.0... 2.00 mg/l
0.0... 5.00 mg/l
0.0...10.00 mg/l
0.0...20.00 mg/l
others
ClO2
1.00 mg/l
end meas. range
temperature
°C Celsius
°F Fahrenheit
temperature
0.0... 50.0 °C
0.0...100.0 °C
-5.0...120.0 °C
pH
0.00...14.00 pH
2.00...12.00 pH
5.00... 9.00 pH
others
pH pH
0.00 pH 14.00 pH
start meas range end meas. range
current output
0...20 mA
4...20 mA
others
current output
0.00 mg/l = 0 mA
start meas range
current output
1.00 mg/l= 20 mA
end meas. range
ClO2 alarm
upper limit
lower limit
hysteresis
alarm delay
water meter
pulse signal
current signal
volume per pulse
1.0 I
max. flow
10 m3/h
current signal
0...20 mA
4...20 mA
others