2.06 Important before using the Scale module and Class III scale
Important/warning
• Read the user manual carefully before using any weighing instrument for the
device.
Safety instructions
• Please comply with the maximum capacity of the scale which is stated on the
identication label on the back of the device.
The maximum load must not be exceeded. The safety requirements and
the notes on appropriate use must also be observed.
• Any modication exempts the supplier from liability for any damage as a
result of modication.
• It is strictly forbidden to carry out any repairs and soldering work on the
motherboards or to replace any components. Repairs must only be
undertaken by persons who are authorized and qualied to do so.
The following applies to Class III scales:
• Sealing label (communication module) may not be broken
• The scale must be re-veried by a Notied Body, cf. local legislation for
maintaining medical approval. If necessary, a Guldmann-approved service
technician can calibrate the scale in connection with a Notied Body’s re-
verication.
• When ordering a Class III scale, indicate the exact operating address (zip code,
city, country). The specic G factor at the place of operation is coded into the
scale’s software, and the scale may only be used at this exact location.
Shock eects
The GH3+ with scale module includes high-sensitive sensors to register mass
corresponding to the nominal recommended load. The sensors are highly
sensitive and can be damaged by the eects of shock, for example pulling
the GH3+ at extreme speed into an end stop.
Operation
Always reset the GH3+ scale module before weighing takes place. When
resetting the unit, the lifting hanger and the desired lifting sling must be
attached under the hoist.
Never pull the hand control wire while weighing.