43
© Guldmann GB/US-02/2021 • # 550216_19
© Guldmann GB/US-02/2021 • # 550216_19
2.13 Safety functions
The emergency stop and emergency lower should only be used in the event
of hoist failure or fault. The fault must be identied and rectied by a suitably
qualied technician before the hoist can be taken back into use. The Emer-
gency stop should not be reset unless by a suitably qualied technician.
Under no circumstances should the user attempt to reset and continue to use
the hoist where the emergency stop has been activated due to fault or failure.
The hoist issue should be referred for service and rectication by a suitably
qualied technician.
Emergency stop and emergency lowering strap
The red strap has the following functions:
• One pull: Emergency stop is activated.
• Constant pull: Emergency lowering is activated.
Emergency stop
If the GH3 does not stop/react to the hand control when the GH3 is in
use, pull the red strap and the lifting/lowering functions (except emergency
lowering) are deactivated. When the emergency stop is activated, the hoist
will not function. The green lamp is switched o.
Reset emergency stop
Reset the emergency stop by pressing the
yellow button on the bottom of the hoist.
The yellow button that appears when the
emergency stop is pulled, must be pressed
manually before the GH3 is ready for use.