GDM-906X Series User Manual
154
Frequency
Calibration
Frequency
Compensate
(1.000017)
Enable or disable frequency compensation
(the value indicates the compensation
coefficient; default: Factory calibration value)
Check the box to enable:
Frequency = Original Frequency x
Compensate Coefficient
Uncheck the box to disable:
Frequency = Original Frequency
FREQ Cali
Mathod
Select either Auto or Manual frequency cali
method. When Manual is opted, input a
frequency compensation coefficient directly.
1. Select Manual mode.
2. Input a compensation coefficient.
3. Use the Left/Right keys to move the
cursor followed by pressing the F5
(Enter) key to save the frequency
compensation coefficient. The value
changes as the figure shown below.
w ww. . com
information@itm.com1.800.561.8187

Table of Contents