REMOTE CONTROL
275
CALCulate:LIMit:LOWer[:DATA]
Sets the lower limit value of the compare function.
Parameter: <NRf> (-1.2E+08 ~ 1.2E+08) | MIN | MAX | DEF
Example: CALC:LIM:LOW:DATA -1.0
Sets the lower limit to -1.0
CALCulate:LIMit:LOWer[:DATA]?
Returns the lower limit value of the compare function.
Return parameter: <NRf>
CALCulate:LIMit:UPPer[:DATA]
Sets the upper limit value of the compare function.
Parameter: <NRf> (-1.2E+08 ~ 1.2E+08) | MIN | MAX | DEF
Example: CALC:LIM:UPP:DATA 1.0
Sets the upper limit to 1.0
CALCulate:LIMit:UPPer[:DATA]?
Returns the upper limit value of the compare function.
Return parameter: <NRf>
CALCulate:LIMit[:STATe]
Sets the status on/off for the compare function.
Parameter: 0 | 1 | ON | OFF
Example: CALC:LIM:STAT 1
Sets the compare function to on.
CALCulate:LIMit[:STATe]?
Returns the status of the compare function.
CALCulate:DB:REFerence
Sets the reference value for the dB function.
Parameter: <NRf> | MIN | MAX | DEF
RefMethod:
Voltage: (-1200 ~ 1200 V)
dBm: (-200.0 ~ 200 dBm)
Example: CALC:DB:REF MAX
Sets the reference value for dB measurements to the maximum allowed.
CALCulate:DB:REFerence?
Returns the reference value from the dB function.
Return parameter: <NRf>
CALCulate:DB:REFerence:METHod
Sets the unit of reference value for the dB function.
Parameter: VOLTage | DBM
Example: CALC:DB:REF:METH DBM
Sets the unit to dbm of reference value for dB function.