GDM-906X Series User Manual
294
[SENSe:]CAPacitance:RANGe:AUTO?
Returns the capacitance Auto-range settings.
Return parameter: 0 | 1, 0=OFF, 1=ON
SENSe CONTinuity Commands
[SENSe:]CONTinuity:NPLCycles
Sets the integration time for Continuity measurement in PLCs (power line
cycles). Where one PLC is equal to 16.6 milliseconds. For any <NRf>
parameter, the DMM will automatically set the PLC to the closest
acceptable PLC value (0.15 | 0.6 | 1).
Parameter: NPLCycles(<NRf> | MIN | MAX | DEF)
Example: SENS:CONT:NPLC MIN
Sets the integration time to the 0.15 PLCs for continuity measurement.
[SENSe:]CONTinuity:NPLCycles?
Returns the integration time for Continuity measurement in PLCs (power
line cycles). Where one PLC is equal to 16.6 milliseconds.
Return parameter: 0.15 | 0.6 | 1
[SENSe:]CONTinuity:RESolution
Sets the Continuity measurement resolution. The resolution depends on the
rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example: SENS:CONT:RES 0.001
Sets the Continuity resolution to 0.001
[SENSe:]CONTinuity:RESolution?
Returns the Continuity measurement resolution.
[SENSe:]CONTinuity:THReshold
Sets the continuity threshold value in ohms.
Parameter: <NR1> (1 ~ 1000)
Example: SENS:CONT:THR 10
Sets the continuity threshold value to 10
[SENSe:]CONTinuity:THReshold?
Returns the continuity threshold value.
Return parameter: <NR1>, Ex: +0010
[SENSe:]CONTinuity:TRIGger:DELay
Sets the trigger delay that minimum step is microseconds of Continuity
measurement.
Parameter: <NRf>(0 ~ 3600 s) | MIN | MAX | DEF
Example: SENS:CONT:TRIG:DEL 0.0001
Sets the trigger delay time to 100us of Continuity measurement.