REMOTE CONTROL
301
[SENSe:]CURRent[:DC]:NULL:VALue:AUTO
Sets the relative value auto on/off for DC Current measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:CURR:DC:NULL:STAT ON
SENS:CURR:DC:NULL:VAL:AUTO ON
The unit automatically sets the 1st count of measurement as null value.
[SENSe:]CURRent[:DC]:NULL:VALueAUTO?
Returns the null value auto state of DC Current measurement.
[SENSe:]CURRent[:DC]:RANGe
Sets the DC current measurement range.
Parameter: Range(<NRf> | MIN | MAX | DEF)
Example: SENS:CURR:DC:RANG 10e-2
Sets the DC current range to 100mA.
[SENSe:]CURRent[:DC]:RANGe?
Returns the DC current measurement range.
[SENSe:]CURRent[:DC]:RANGe:AUTO
Sets the DC current Auto-range settings on, off or once.
Parameter: 0 | 1 | ON | OFF | ONCE
Example: SENS:CURR:DC:RANG:AUTO OFF
Turns Auto-range off for DC current measurement.
[SENSe:]CURRent[:DC]:RANGe:AUTO?
Returns the DC current Auto-range settings.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]CURRent[:DC]:RESolution
Sets the DC Current measurement resolution. The resolution depends on
the rate and range settings.
Parameter: Resolution(<NRf> | MIN | MAX | DEF)
Example: SENS:CURR:DC:RES 0.01
Sets the DC Current resolution to 0.01
[SENSe:]CURRent[:DC]:RESolution?
Returns the DC Current resolution.
[SENSe:]CURRent[:DC]:TERMinals
Assigns an input port for the current function.
Parameter: <NR1> GDM-9060 : 3 / GDM-9061 : 3 | 10
Example: SENS:CURR:DC:TERM 3
Sets the input jack to the 3A current input port.
[SENSe:]CURRent[:DC]:TERMinals?
Returns the assigned input port used for the current function.
Return parameter: +3 | +10