REMOTE CONTROL
309
[SENSe:]FREQuency:INPutjack
Assigns an input port for the frequency function.
Parameter: <NR1> (0 | 1 | 2), 0=Voltage, 1=3A, 2=10A
Example: SENS:FREQ:INP 0
Sets the input jack to the Voltage input port.
[SENSe:]FREQuency:INPutjack?
Returns the assigned input port used for the frequency function.
Return Parameter: VOLT | 3A | 10A
[SENSe:]FREQuency:NULL[:STATe]
Sets the relative function on/off for Frequency measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:FREQ:NULL:STAT ON
Turns the relative function on for Frequency measurement.
[SENSe:]FREQuency:NULL[:STATe]?
Returns the relative function state of Frequency measurement.
Return parameter: 0 | 1, 0=OFF, 1=ON
[SENSe:]FREQuency:NULL:VALue
Sets the relative value for Frequency measurement.
Parameter: <NRf> (-1.2e6~1.2e6 Hz) | MIN | MAX | DEF
Example: SENS:FREQ:NULL:VAL 10
Sets the relative value to 10Hz for Frequency measurement.
[SENSe:]FREQuency:NULL:VALue?
Returns the current relative value of Frequency measurement.
[SENSe:]FREQuency:NULL:VALue:AUTO
Sets the relative value auto on/off for Frequency measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:FREQ:NULL:STAT ON
SENS:FREQ:NULL:VAL:AUTO ON
The unit automatically sets the 1st count of measurement as null value.
[SENSe:]FREQuency:NULL:VALue:AUTO?
Returns the null value auto state of Frequency measurement.
[SENSe:]FREQuency:TIMeout:AUTO
Assigns timeout time at the frequency measurement.
Parameter: 0 | 1 | ON | OFF
Example: SENS:FREQ:TIM:AUTO OFF
Sets the timeout time at 1 seconds.