GDM-906X Series User Manual
338
*PSC?
Returns power on clear status.
Return parameter: <Boolean>(0|1) 0= disables, 1= enables
*RCL
Load the system parameters from 1 of 5 memory locations.
Parameter: <NR1> (0~4) (1~5= memory number)
Example: *RCL 1
Loads the memory 2 system parameters.
Note: prior to loading memory, the corresponding memory data is required;
otherwise, invalid data will be loaded and therefore command error will be shown.
*RST
Recalls default panel setup.
Resets instrument to factory default state. This is similar to
SYSTem:PRESet. The difference is that *RST resets the instrument for SCPI
operation, and SYSTem:PRESet resets the instrument for front panel
operation. As a result, *RST turns the histogram and statistics off, and
SYSTem:PRESet turns them on.
*SAV
Save the system parameters to 1 of 5 memory locations.
Parameter: <NR1> (0~4) (1~5= memory number)
Example: *SAV 2
Saves the system parameters to memory 3.
*SRE?
Returns the SRER (Service Request Enable Register) contents.
*SRE
Sets SRER contents.
Parameter: <NR1>(0~255)
Example: *SRE 7
Sets the SRER to 00000111.
An enable register defines which bits in the event register will be reported
to the Status Byte register group. You can write to or read from an enable
register.