REMOTE CONTROL
137
Queries the Load CR level.
: LOAD2: RESistor?
Return the Load CR level for CH2.
:LOAD:DISPlay {Low|Middle|High}
In load mode, set the minimum voltage value that
can be displayed. The factory default is high.
Below this value, it will display --.----
Low: 0.5V;Middle: 0.75V;High:1V
:LOAD:DISPlay Low
In load mode, the minimum voltage value that can
be displayed is 0.5V.
Query the minimum voltage value that can be
displayed in load mode.
:LOAD:DISPlay?
In load mode, query the minimum voltage value
that can be displayed. The return variables are low,
0.5V; middle, 0.75V; high, 1V.
:DELAy[1|2]:CYCLEs {N|I}[,<value>]
Sets the cycle number of Delay.
<value> 1-99999
I represents an infinite loop. N represents a finite
loop. The cycle number is set by parameter <value>
:DELAy2:CYCLEs N,100
Sets the cycle number of Delay as 100 for CH2.