REMOTE CONTROL
159
:SEQUence[1|2]:TEMPlet:FALLRate <value>
Sets the falling degree of ExpFall.
<value> Integer from 0 to 10.
:SEQUence2:TEMPlet:FALLR 5
Sets the CH2 falling degree of ExpFall as 5.
:SEQUence[1|2]:TEMPlet:FALLRate?
Queries the falling degree of the active ExpFall
:SEQUence2:TEMPlet:FALLRate?
Queries the CH2 falling degree of the active
ExpFall. Returns to an integer ranging from 0 to 10.
:SEQUence[1|2]:TEMPlet:INTErval <value>
<value> integer 1s to 300s
Time interval is the duration period in accord with
each time parameter of the active selected templet.
:SEQUence2:TEMPlet:INTErval 15
Sets the CH2 time interval as 15s. Pulse templet
does Not support this parameter.
:SEQUence[1|2]:TEMPlet:INTErval?
Queries the active time interval.
:SEQUence2:TEMPlet:INTErval?
Queries the CH2 active time interval. Returns to an
integer between 1 to 300.
:SEQUence[1|2]:TEMPlet:INVErt {ON|OFF}
Enable or disable invert function of the active
selected templet.