Setup
23
AC power cable requirement
Here is the AC power cable specification, in case of using cables
other than the attached one.
Cable length
3m
Cable gland
KSS or PG-2013
Cable type
(recommended)
Model: SJT
Type: 3 x 14 AWG stranded copper
Rating: 60°C min, 300V
Diameter: 9.143~10.03 mm
Model: H05 VV-F
Type: 3G 1.5mm
2
stranded copper
Rating: 300V/500V
Diameter: 8.5 ± 0.2 mm
PSH User Manual
24
Remote Sensing and Local Sensing
Remote sensing compensates the cable loss between PSH and load,
up to 0.5V. Use remote sensing whenever the load voltage has to
be accurate.
Local Sensing
(default)
VSET = VOUT
V
OUT > VLOAD
VOUT VLOAD
VOUT
Output
Setting
VSET
|VSET–VOUT|
Sense
PSH Load
The sense terminal is internally connected to the
PSH output terminal. The delta between the
voltage setting level (V
SET) and the actual output
level (V
OUT) is compensated. The load terminal
voltage (V
LOAD) might become lower than the
output due to cable loss.
Remote
Sensing
VSET = VLOAD
VOUT > VSET
PSH Load
Sense
V
OUT VLOAD
VLOAD
Output
Setting
VSET
|VSET–VLOAD|
The sense terminal is connected to the load input
terminal. The delta between the voltage setting
level (V
SET) and the actual load voltage (VLOAD) is
compensated. The output voltage (V
OUT) might
become higher than the setting due to the
compensation.