PSU Series Programming Manual 
114 
External control is not performed. 
Returns output status of the instrument. 
SYST:CONF:OUTP:EXT 1 
SYST:REB  
Active the external output control and reboot the 
unit to active the setting. 
:SYSTem:CONFigure:MONitor:RANGe 
 
This command is used to select the monitor 
voltage range. This is the equivalent to the F-96 
(Monitor Voltage Select) power on configuration 
settings. 
The setting will only be valid after the power has been 
cycled. 
:SYSTem:CONFigure:MONitor:RANGe 
{<NR1>|LOW|HIGH} 
:SYSTem:CONFigure:MONitor:RANGe? 
Returns the range setting. 
SYST:CONF:MON:RANG LOW 
SYST:REB  
Set monitor voltage range to 0~5V and reboot the 
unit to active the setting. 
:SYSTem:CONFigure:CONTrol:RANGe 
 
This command is used to select the external 
analog control voltage (or resistance) range. This 
is the equivalent to the F-97 (Control Range) 
power on configuration settings.