PSU Series Programming Manual
78
Initiate Commands
:INITiate:CONTinuous[:TRANsient] ............................... 78
:INITiate[:IMMediate]:NAME ........................................... 78
:INITiate[:IMMediate][:TRANsient] ................................. 79
:INITiate:CONTinuous[:TRANsient]
Set
Query
Description
This command continuously initiates software
triggers for the transient or output triggers.
Syntax
Query Syntax
:INITiate:CONTinuous[:TRANsient]
{<bool>|OFF|ON}
:INITiate:CONTinuous[:TRANsient]?
Parameter
OFF | 0
OFF
ON | 1
ON
Return parameter
0
OFF
1
ON
Example
INIT:TRAN 1
Turns on the continuous trigger.
:INITiate[:IMMediate]:NAME
Set
Description
The INITiate command starts the TRANsient or
OUTPut trigger.
Syntax
:INITiate[:IMMediate]:NAME {TRANsient|OUTPut}
Parameter
TRANSient
Starts the TRANsient trigger.
OUTPut
Starts the OUTPut trigger.
Example
INITiate:NAME TRANient
Starts the TRANSient trigger.