PSU Series Programming Manual
86
:OUTPut:MODE
Set
Query
Description
Sets the PSU output mode. This is the equivalent
to the F-03 (V-I Mode Slew Rate Select) settings.
Syntax
Return Syntax
:OUTPut:MODE {<NR1>|CVHS|CCHS|CVLS|CCLS}
:OUTPut:MODE?
Parameter
CVHS | 0
CV high speed priority
CCHS | 1
CC high speed priority
CVLS | 2
CV slew rate priority
CCLS | 3
CC slew rate priority
Return parameter
<NR1>
Returns the output mode.
:OUTPut[:STATe][:IMMediate]
Set
Query
Description
Turns the output on or off.
Syntax
Query Syntax
:OUTPut[:STATe][:IMMediate] { <bool> | OFF | ON }
:OUTPut[:STATe][:IMMediate]?
Parameter
OFF | 0
Turns the output off.
ON | 1
Turns the output on.
Return parameter
<bool>
Returns output status of the instrument.
:OUTPut[:STATe]:TRIGgered
Set
Query
Description
Turns the output on or off when a software trigger
is generated.
Syntax
Query Syntax
:OUTPut[:STATe]:TRIGgered { <bool>|OFF|ON }
:OUTPut[:STATe]:TRIGgered?
Parameter
OFF | 0
Turns the output off when a software
trigger is generated (*TRG).