SFG-2000 Series User Manual
30
Set Offset
SFG can add or delete offset to the sine/square/triangle
waveform, thus changing the waveform vertical position.
Use the OFFSET knob.
Activate
offset
Pushed: Offset Off
Pulled: Offset On
Adjust offset
CAL
ADJ
OFFSET
-
+
Turn the knob right (higher position)
or left (lower position).
The range is 5V ~ +5V for 50
load.
Limitation
Note that the output amplitude, including the offset, is
still limited to:
5 ~ +5V (50 load)
10 ~ +10V (no load)
Therefore excessive offset leads to peak clip as below.
Positive peak
clip (50)
+5V
-5V
Offset
Clipped
Negative peak
clip (50)
+5V
-5V
Offset
Clipped