APPLICATION EXAMPLES
47
Test step
1. Apply a triangle waveform to the amplifier first.
Adjust the waveform amplitude to make sure there is
no clipping.
2. Switch to square waveform and adjust its frequency
to the middle of the amplifier pass band, such as
20Hz, 1kHz, and 10kHz.
3. Observe the shape of the amplifier output. The
following table shows the possible output distortions
and their explanations.
Amplitude reduction at low
frequency
No phase shift
Transient
characteristic
list
Low frequency boosted (accentuated
fundamental)
High frequency loss
No phase shift
Low frequency phase shift
Trace thickened by hum-voltage
High frequency loss
Phase shift
Low frequency loss
Phase shift
Low frequency loss
Low frequency phase shift
High frequency loss
Low frequency phase shift
Damped oscillation
Note
For narrow band amplifier testing, square wave may not
be suitable.