8.3 Set the sound options
The meter can make an audible sound when a key is pushed, when a
reading is complete or when the calibration reminder is due.
1. Push SETTINGS and select Sounds.
2. Select which events will produce an audible sound. Multiple items
can be selected.
Option Description
Key Press The meter will make an audible sound
whenever a key is pushed.
Reading
complete
The meter will make an audible sound
whenever a reading is completed.
Reminders The meter will make an audible sound when a
calibration is due.
8.4 Security options
The Security Options menu is used to protect the meter setup.
The Setup Date and Time, Delete Data Log, Restoring Factory Defaults
and Restore Factory Cal screens are not accessible without a password.
Store the password in a safe and accessible place. If the specified
password is forgotten and Security Options is turned on, the operator is
locked out of the restricted menus. Contact technical support if the
password is lost.
8.4.1 Turn security options on
The security options and the set password options are used together to
prevent access to restricted menus.
1. Push the SETTINGS key and select Security Options.
2. Select Edit Password and use the UP and DOWN keys to set a
password.
3. Select Security On to enable the password setting. The requirement
for the password entry is controlled by setting Security Options on or
off.
Note: Set the Security to Off to disable the password setting.
4. Push the ON/OFF key to turn off and on the meter to activate the
password settings.
8.5 View meter information
The instrument information menu shows specific information such as the
meter name, model number, software version, serial number and
available Operator and Sample IDs. 10 Operator IDs and 100 Sample
IDs are available.
1. Push the SETTINGS key and select Meter Information.
8.6 Calibration
The portable turbidimeter is calibrated with Formazin Primary Standards
at the factory. The meter should be calibrated upon receipt for best
results. The manufacturer recommends calibration with a primary
standard such as StablCal Stabilized Standards or with formazin
standards every three months.
Note: Set Cal Reminder Repeat in the Calibration Options menu for periodical
calibration. Verify the calibration once a week.
8.6.1 Calibration options
The calibration options contain Calibration History, Calibration Curves,
Cal Reminder Repeat and Restore Factory Calibration.
1. Push the CALIBRATION key and then the UPand DOWN key.
Option Description
Calibration
History
The calibration history shows a list of the times
when the meter was calibrated. Select a date
and time to view a summary of the calibration
data.
14 English

Other manuals for Hach 2100Q

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100Q and is the answer not in the manual?

Hach 2100Q Specifications

General IconGeneral
Range0 - 1000 NTU
Resolution0.01 NTU on lowest range
DetectorSilicon photodiode
Stray Light< 0.02 NTU
Measurement MethodNephelometry
Power Supply4 AA alkaline batteries
Operating Temperature0 to 50 °C
Storage Temperature-20-60°C
Measurement ModesSignal Averaging, Rapidly Settling Turbidity
DisplayLCD
Accuracy±2% of reading plus stray light from 0 to 1000 NTU
Light SourceTungsten filament lamp