Error/Warning Description Solution
Error - Security
Please set password
before activating
security
No password is
created.
Create a new
password.
Please enter at least
one character.
Password must
contain minimum of
one character.
Create a password of
at least one character.
Password incorrect.
Please retry.
Incorrect password
was entered.
Enter the appropriate
password.
Please disconnect
the USB cable from
your computer.
Data storage does not
respond while
connected to the
meter and the
computer.
Disconnect the USB
cable from the meter
and try sending data
again.
USB module
memory full. Delete
data and try again.
Data storage is full. 1. Connect
USB/power module to
the computer.
2. Download the
stored data to the
computer.
3. Delete Data Log on
the module.
Error/Warning Description Solution
Delete Last Reading
Failed!
Error in the data
storage.
Turn the meter off and
on. If the error
message still occurs,
contact technical
support.
Delete Data Log
failed!
Can't read data set!
Can't store data!
Can't store to the
Reading Log!
Can't store to the
Verify Cal Log!
Error storing data!
Error reading data!
Section 11 Replacement parts and accessories
11.1 Replacement parts
Description Quantity Item no.
StablCal ampule calibration kit 1 2971205
10 NTU verification standard 100 mL 2961701
Silicone Oil 15 mL 126936
Insert, molded bottom 1 2971507
Sample cell oiling cloth 1 4707600
1" glass sample cell (10 ml ) w/cap
(Turb)
pkg/6 2434706
Carrying case (includes insert) 1 2971500
26 English