8.9 Indexing a single cell
Precise measurements for very low turbidity samples require the use of
a single cell for all measurements or optically matching the cells. Use
one cell to provide the best precision and repeatability. When one cell is
used, an index or orientation mark (other than the factory-placed
diamond) can be placed on the cell so it is inserted into the instrument
with the same orientation each time.
When using a single cell, make an index or orientation mark on the cell
as follows:
20 English

Other manuals for Hach 2100Q

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100Q and is the answer not in the manual?

Hach 2100Q Specifications

General IconGeneral
Range0 - 1000 NTU
Resolution0.01 NTU on lowest range
DetectorSilicon photodiode
Stray Light< 0.02 NTU
Measurement MethodNephelometry
Power Supply4 AA alkaline batteries
Operating Temperature0 to 50 °C
Storage Temperature-20-60°C
Measurement ModesSignal Averaging, Rapidly Settling Turbidity
DisplayLCD
Accuracy±2% of reading plus stray light from 0 to 1000 NTU
Light SourceTungsten filament lamp