8.6.3 StablCal RapidCal calibration
1. Push the
CALIBRATION key
to enter the
Calibration mode.
Follow the
instructions on the
display.
Note: Gently invert
each standard before
inserting the standard.
2. Push the UP and
DOWN key to
access Cal Options
and then select
Cal.Curve.
3. Select StablCal
RapidCal from the
list and push OK.
4. Insert the 20 NTU
StablCal Standard
and close the lid.
Note: The standard to
be inserted is bordered.
5. Push Read. The
display shows
Stabilizing and then
shows the result.
6. Push Done to
review the
calibration details.
16 English