EasyManuals Logo

Hach 2100Q User Manual

Hach 2100Q
34 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
Figure 2 2100Q and 2100Qis components
1 2100Q or 2100Qis turbidimeter 6 Silicone oil
2 Carrying case 7 20, 100 and 800 NTU StablCal
calibration standards
3 User manual and Quick
reference guide
8 AA alkaline batteries (pk/4)
4 Oiling cloth 9 StablCal 10 NTU verification
standard
5 1" sample cell (10 mL) with
cap (pk/6)
Section 3 Installation
C A U T I O N
Multiple hazards. Only qualified personnel must conduct the
tasks described in this section of the document.
3.1 Install the battery
W A R N I N G
Explosion hazard. An expired battery can cause hydrogen
gas buildup inside the instrument. Replace the battery
before it expires. Do not store the instrument for long
periods with a battery installed.
W A R N I N G
Potential fire hazard. Use only alkaline or nickel metal hydride batteries
(NiMH) in the meter. Other battery types or incorrect installation can
cause a fire. Never mix battery types in the meter.
N O T I C E
The battery compartment is not waterproof. If the battery compartment
becomes wet, remove and dry the batteries and dry the interior of the
compartment. Check the battery contacts for corrosion and clean them
if necessary.
N O T I C E
When using nickel metal hydride (NiMH) batteries, the battery icon will
not indicate a full charge after freshly charged batteries have been
inserted (NiMH batteries are 1.2 V versus 1.5 V for alkaline batteries).
Even though the icon does not indicate complete charge, 2300 mAH
NiMH batteries will achieve 90% of instrument operation lifetime
(before recharge) versus new alkaline batteries.
N O T I C E
To avoid potential damage to the meter from battery leakage, remove
the meter batteries prior to extended periods of non-use.
6 English

Other manuals for Hach 2100Q

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach 2100Q and is the answer not in the manual?

Hach 2100Q Specifications

General IconGeneral
Range0 - 1000 NTU
Resolution0.01 NTU on lowest range
DetectorSilicon photodiode
Stray Light< 0.02 NTU
Measurement MethodNephelometry
Power Supply4 AA alkaline batteries
Operating Temperature0 to 50 °C
Storage Temperature-20-60°C
Measurement ModesSignal Averaging, Rapidly Settling Turbidity
DisplayLCD
Accuracy±2% of reading plus stray light from 0 to 1000 NTU
Light SourceTungsten filament lamp

Related product manuals