Specification Details
Vertical sample lift 8.5 m (28 ft) for 8.8 m (29 ft) maximum of
3
/
8
-in. vinyl intake tube at
sea level at 20 to 25 °C (68 to 77 °F)
Pump flow rate 4.8 L/min (1.25 gpm) at 1 m (3 ft) vertical lift with
3
/
8
-in. intake tube
typical
Sample volume Programmable in 10-mL (0.34 oz) increments from 10 to 10,000 mL
(3.38 oz to 2.6 gal)
Sample volume repeatability
(typical)
±5% of 200 mL sample volume with: 4.6 m (15 ft) vertical lift, 4.9 m
(16 ft) of
3
/
8
-in. vinyl intake tube, single bottle, full bottle shut-off at
room temperature and 1524 m (5000 ft) elevation
Sample volume accuracy
(typical)
±5% of 200 mL sample volume with: 4.6 m (15 ft) vertical lift, 4.9 m
(16 ft) of
3
/
8
-in. vinyl intake tube, single bottle, full bottle shut-off at
room temperature and 1524 m (5000 ft) elevation
Sampling modes Pacing: Fixed Time, Fixed Flow, Variable Time, Variable Flow,
Event
Distribution: Samples per bottle, bottles per sample and time based
(switching)
Run modes Continuous or non-continuous
Transfer velocity (typical) 0.9 m/s (2.9 ft/s) with: 4.6 m (15 ft) vertical lift, 4.9 m (16 ft) of
3
/
8
-in.
vinyl intake tubing, 21 °C (70 °F) and 1524 m (5000 ft) elevation
Liquid detector Ultrasonic. Body: Ultem
®
NSF ANSI standard 51 approved, USP
Class VI compliant. Contacting liquid detector or optional non-
contact liquid detector
Air purge An air purge is done automatically before and after each sample.
The sampler automatically compensates for different intake tube
lengths.
Tubing Intake tubing: 1.0 to 30.0 m (3.0 to 99 ft) length, ¼-in. or
3
/
8
-in. ID
vinyl or
3
/
8
-in. ID Teflon
™
-lined polyethylene with protective outer
cover (black or clear)
Wetted materials Stainless steel, polyethylene, Teflon, Ultem, silicone
Memory Sample history: 4000 records; Data log: 325,000 records; Event log:
2000 records
Communications USB and optional RS485 (Modbus)
Electrical connections Power, auxiliary, optional sensors (2x), USB, distributor arm,
optional rain gauge
Analog outputs AUX port: none; optional IO9000 module: Three 0/4–20 mA
outputs to supply the recorded measurements (e.g., level, velocity,
flow and pH) to external instruments
Analog inputs AUX port: One 0/4–20 mA input for flow pacing; optional
IO9000 module: Two 0/4–20 mA inputs to receive measurements
from external instruments (e.g., third-party ultrasonic level)
Digital outputs AUX port: none; optional IO9000 module: Four low voltage,
contact closure outputs that each supply a digital signal for an alarm
event
English 5