Page 105
8.2.5 New Reagents Program
SPAN CALIBRATION (NO by default) defines if the Span
Calibration reactions are part of the Install New Reagents
procedure in Reagents Setup menu.
If SPAN CALIBRATION is activated (YES), the standard solution
needs to be connected to the Calibration/Manual port of the
BioTector as part of the Install New Reagents procedure. See
concentrations of the TOC STANDARD defined in Span
Calibration menu (see Section 2.3.2 Span Calibration
for details).
SPAN CHECK (NO by default) defines if the Span Check
reactions are part of the Install New Reagents procedure in
Reagents Setup menu.
Note that both SPAN CALIBRATION and SPAN CHECK
functions cannot be selected as YES simultaneously and cannot
occur together in the Install New Reagents procedure.
AUTOMATIC RE-START (Yes by default) defines BioTector to
return automatically online operation having completed the
Install New Reagents cycle.