EasyManua.ls Logo

Hach BioTector B7000i - Section 6 Reagents and Calibration Standards; 6.1 Reagents

Hach BioTector B7000i
171 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 64
Section 6 Reagents and Calibration Standards
6.1 Reagents
BioTector TOC analyzer uses following reagents:
I. Acid: 1.8 N Sulfuric Acid (H
2
SO
4
) Reagent containing 80mg/l Manganese Sulfate Monohydrate
II. Base: 1.2 N Sodium Hydroxide (NaOH) Reagent
Reagents should not contain high levels of organics, nitrates and phosphates. Ideally, the level of organics,
nitrates and phosphates should be less than 100 g/l (ppb) in the deionized water used to prepare TOC
analyzer reagents.
Acid and Base reagents are stable up to 1 year. Table 7 below summarizes the total days each BioTector
TOC Analyzer reagent lasts at various system analysis range configurations:
Table 7 BioTector TOC Analyzer Reagent Consumption
REAGENTS
Container Size
(Liters)
TOTAL DAYS REAGENT LASTS PER ANALYSIS
RANGE
0-250
mgC/l
0-2000
mgC/l
0-20000
mgC/l
Acid
25
54
34
32
Base
25
53
33
31
Above table is derived from several operation parameters such as 100% online time. Recommended bunds
(reagent spill trays) to contain above quantity reagents are 1x 50 Liters.