Page 129
OUTPUT 1-8
OUTPUT1-8 menus contain system optional output relay
settings. Relay terminals 1-8 are located on the stream
expansion board (auxiliary board), which is an optional feature of
the system.
OUTPUT1-8 relays can be programmed to a single output
function or to multiple output functions as described for RELAY
17-20 above. STREAM 1 is always functional as default within
the system software.
The programmed output function or functions are marked with
an asterisk “*” sign in these menus as in the example for
OUTPUT 2, which is programmed for STREAM 2 below if this
option is installed into the BioTector.
When multiple functions are programmed for an output relay, the
relay output is activated when one or multiple conditions are
initiated.
Stream and manual grab sample selection outputs are assigned
to one output function only. In the example below, OUTPUT 2 is
activated to select stream channel 2 function only. Any output
menu, assigned for a specific stream or for a grab sample
channel output function, should not be programmed for multiple
output functions.
In BioTector configuration data download (or in All Data
download), the programmed output functions for each output is
tabulated and marked with asterisk “*” signs for clarity as
illustrated below.
OUTPUT 2
DEFAULT STATE N/E
……………………… ..
CAL SIGNAL
STREAM 1
STREAM 2 *
STREAM 3
……………………… ..