Figure 8 below illustrates the BioTector door clearance dimensions.
Figure 8 BioTector Door Clearance Dimensions
When BioTector is being mounted on a wall or a stand, the support has to be strong enough to carry
typically four times of the weight of BioTector (~400 kg).
The BioTector should be lifted applying a safe method in accordance with local regulations.
The minimum size of the bolts used to hold the BioTector in place should be M8.