118
Expanded programs
10. Follow the directions on the display and set various filters
one after another in the order displayed.
11. Press Next after inserting a filter.
After the last reading the results are displayed.
12. Press PC & Printer icon to send the data to a USB memory
stick, PC or to a printer.
The files will be stored automatically as CSV files (Comma
Separated Value). The file name will be formatted as
"Verification.csv".
6.7.3.2 Wavelength check
The Wavelength Check test is used to check wavelength
accuracy at 656.1 nm and 486.0 nm. The bandwidth is checked
at 656.1 nm.
1. Select Wavelength Check from the "Optical Checks" menu.
2. Press to select the wavelength.
3. Press OK to confirm.
4. Press Start.
Note: During the UV lamp warm-up phase, the text "Lamp
Warmup"is shown.
The result is shown.
5. Press Cancel to go back to the Optical Checks.