Standard programs
5.2.2.3 Import external sample ID
1. Create an external sample ID list with a spreadsheet
program. Four columns are permitted, column 1 contains the
sequential number, column 2 the sample designation,
columns 3 and 4 are optional. Headings and comment lines
must begin with #.
2. Create a folder named "SampleID" on a USB stick or in your
established network.
3. Save the sample ID list with Save As in the format CSV or
UNICODE-TXT in the folder "SampleID".
4. Connect the DR 6000 to the USB stick or network.
All *.TXT and *.CSV files in the SampleID folder are shown
for selection.
5. Transfer the shown sample ID list with Done.
6. Select the required file with OK.
5.2.3 Sample ID and operator ID with the optional RFID Sample ID Kit (not available on
all models)
The optional RFID Sample ID Kit consists of:
A LOC 100 RFID Locator
An operator RFID tag
Five location RFID tags and
Three sample bottle RFID labels each, in black, red, green,
blue and yellow
The operator and location RFID tags can be written to and read in
by the RFID module in the DR 6000 as often as required.
More information can be found in the user manual for the
LOC 100 RFID Locator.
5.2.4 Security settings
The "Security" menu contains a variety of security settings to
control access to various functions.
All functions can be allocated to three different security levels:
not secured: every operator can make changes in this area.
one key: every operator assigned this security level can
make changes to non-secured functions and functions with
one key.
two keys: every operator assigned this security level can
make changes in all functions

Table of Contents

Other manuals for Hach DR 6000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hach DR 6000 and is the answer not in the manual?

Hach DR 6000 Specifications

General IconGeneral
Wavelength Range190 - 1100 nm
Light SourceTungsten and Deuterium lamps
DetectorSilicon photodiode
Wavelength Accuracy±1 nm
Wavelength Repeatability< 0.1 nm
Photometric Range± 3.0 Abs
Photometric Linearity±0.002 Abs at 0.0 to 1.0 Abs
Stray Light< 0.05%T at 340 nm
Display7 inch color touchscreen
Power Requirements100 - 240 V / 50 - 60 Hz
ConnectivityUSB, Ethernet