68
Standard programs
5.4.11 Analysis of samples
1. Press Stored Programs and select a program.
Note: If available, the procedural instruction is indicated on the
display via the Info icon.
2. Insert the zero solution cell into the cell compartment.
3. Press Zero.
4. Remove the zero solution cell from the cell compartment.
Insert the sample cell into the cell compartment.
5. Press Read. The result will be displayed.
Note: To define a sample dilution, press the Dilution key in the
toolbar.
6. For data storage, refer to section 5.3.1, page 44.
5.4.12 Program tests
5.4.12.1 Program a new test
1. Press Stored Programs in the Main Menu.
2. Press Program Options and then New.
3. Use the alphanumeric keypad to enter the Program Number
indicated in the working procedure.
An overview of the data for the test to be programmed is
displayed.
The process specification of the corresponding test contains
the new data (Wavelength, Factors, reading Ranges,
Conversion Factor for chemical Form etc.).
4. Highlight the line containing the test specification that is to be
programmed, then press Edit and enter the data in the
appropriate program lines.
5. Press OK and then Store after editing the test.
5.4.13 Add stored programs to the favorite programs list
The "Favorites" menu simplifies test selection by creating a list of
the most frequently used tests from the Stored Programs and
User Programs.