20
Installation
3.5 Interfaces
The DR 6000 features three USB ports and one Ethernet port as
standard, located on the front and back of the instrument
(Figure 1 and Figure 2, Page 19).
The USB type A ports are used for communications with a printer,
USB memory stick or keyboard. A USB memory stick can be
used to update the instrument software.
The USB type B port is used for communications with a PC. The
optional Hach Data Trans software (refer to Section 9, Page 143)
must be installed on the PC for data transfer.
A USB hub may be used to connect several accessories at a
time.
Note: USB cables must not be longer than 3m. Use USB cables with
ferrite shielding.
These interfaces provide the output of data and graphics to
printers and PCs and the updating of the device software (refer to
section 6.7.2, page 116). The Ethernet port supports real-time
data transfer in local networks, LIMS systems and SC controllers.
Only use a shielded cable (for example STP, FTP, S/FTP) with a
maximum length of 20 m for the Ethernet port.
3.6 Cell compartments and universal cell adapter
3.6.1 Cell compartments and adapter
The DR 6000 has two cell compartments (Figure 3). Only one cell
type at a time can be used for a reading.
Cell compartment (1) for:
13 mm and 16 mm round cells
Note: Cell compartment (1) contains a barcode detector for
cells/vials.
Cell compartment (2) for:
The following cell types can be inserted into the cell compartment
(2) via the universal cell adapter.
10 mm square cells
20 mm square cells
30 mm square cells
50 mm square cells
1 inch square cells
1 inch round cells
1 inch Pour-Thru cells
AccuVac
®
vials