126
Expanded programs
6.7.4.6 Pipette check
This function and the optional accessories tests the accuracy of
the pipette test set. To carry it out follow the instructions in the
pipette test set.
1. Press Pipette Check.
2. Press Options.
3. Select New.
4. Define the pipette to be checked, the wavelength and the
upper and lower tolerance limits.
5. Select Close.
6. Select the pipette to be checked from the selection list and
press Start.
7. Insert the zero cell and press Zero.
A zero measurement is performed.
8. Insert the analysis cell and press Read.
A measurement is performed.
A test log indicating the target and actual value and whether
or not the test has been passed is shown.
9. Press the PC & Printer icon to send the test log to the printer.
10. Press Close to return to the selection list.