pH Operations
27
7. When calibration is complete for that buffer, the meter will beep three times,
and the display will show a flashing CAL 2 and the next pH buffer value factory
default is 4.01).
8. Rinse the probe in deionized water and place in second buffer solution.
9. Push ENTER if OK, or to change value. Push
PH to accept a 1-point calibration
and return to the pH measurement mode.
10. If ENTER is selected, the buffer value will start flashing until calibration is
complete.
11. If the meter is set for only a two point calibration, upon completion of the
second point the meter will beep three times and the display will return to the
pH measurement mode. If the meter is set up for three point calibration the
display will show a flashing CAL 3 and the next pH value (default is 10.01).