49
(10) Inflate the occluding cuff at the thigh to at least
60 mmHg. Pressures in the cuff are plotted on
the screen. The graph will indicate a gradual
increase in waveform amplitude signifying that
venous outflow is blocked by the occluding cuff.
(11) After 90 seconds, disconnect the sphyg from
the occluding cuff. The pressures on the screen
will drop back to the baseline.
(12) Smartdop will automatically stop the
measurement process, and then freeze the
waveform. Press Print Button to print the
waveform if necessary.
(13) If you wish to save the data on the memory, see
“ 3-2-3-a. MEMORY - STORE”.
(14) Press the Right to get out of the freeze mode.