53
6-3. Specifications
Probes: Model Freq. Ispta* (in situ) [mW/cm²]
BT2M20S8C 2 MHz 80 mW/cm
²
or less
BT4M05S8C 4 MHz 390 mW/cm
²
or less
BT5M05S8C 5 MHz 390 mW/cm
²
or less
BT8M05S8C 8 MHz 390 mW/cm
²
or less
BT10M5S8C 10 MHz 390 mW/cm
²
or less
*Ispta : Spatial Peak - Temporal Average Intensity
Power: Ni-MH rechargeable battery pack or AC adaptor
Input: AC 100-240V- 0.3A (max), 50/60 Hz
Output: DC 12V, 500mA or more
Consumption: DC 12 V, 300 mA
Recharge: Approx. 3 hours by the AC adaptor
Full charge life: 2.5 hours or more if used with freeze mode.
Battery life: Approx. 2 years, 300 full charges
Automatic shut-off No signal: 5 min. Freeze: 10 min.
Others: 15 min. (only FHR WAVE mode: 35 min.)
Frequency range: 80 / 200 Hz to 5 kHz
Mode settings: Memory, Waveform, Direction, Time scale, Others
Waveform memory: 30 waveforms
LCD display: 128 x 64 dots, STN LCD
Bi-directional waveform (normal & slow mode)
Numerical data (Systolic, diastolic & mean velocities, RP, PI, SD,
HR)
Heart rate: 30 to 300 BPM, accuracy of ±5%
Battery level and low battery indicators
Printer: Paper: 58 mm (W) x 25 m/roll (L), Thermal
Resolution: 384 dots/line
Print speed: 25 mm/s
Velocity accuracy: ±10% or less comparing with internal phantom testing.
Speaker output: 250 mW or more