Microlab
®
STAR User Manual
Page 38 of 225 610766/02
1HHGOH:DVK6WDWLRQ
The needle wash station has its own power supply. The mains plug is on the rear of the
pump station (see picture below).
Ensure that the the needle wash station is correctly earthed when connected to the power
supply. Connect the wash station only to an earthed outlet.
Ensure that the voltage selector is correctly set before operating the needle wash station.
The fuses for the instrument are situated next to the mains power switch (see picture below).
The pump station has two fuses for the power supply which can be accessed by opening the
cover above the mains plug.
)XVHVIRUSRZHUVXSSO\RIZDVK VWDWLRQ.
The technical specifications regarding electrical power for the wash station are listed in the
following table:
Maximum power consumption:
Voltage:
Frequency:
Delayed action fuse:
115 V∼:
230 V∼:
140 W
115 / 230 V∼ -15 % + 15 %
50 / 60 Hz ± 5 %
3.15 A (T3.15L250)
1.6 A (T1.6L250)
Overvoltage category II
Degree of pollution 2