Microlab
®
STAR User Manual
610766/02 Page 39 of 225
0DLQWHQDQFH
A short preventive maintenance including volume verification should be carried out twice
yearly. Typically, the service and volume verification are carried out by a trained HAMILTON
service engineer.
To clean the instrument deck use the procedures required by GLP. To preserve plastic
materials from damage, do not use organic solvents.
If a spray is used, do not point the spray directly at the autoload unit or at electrical boards or
connectors.
9HULILFDWLRQ
The Microlab STAR will be verified by a trained HAMILTON technician upon initial set-up,
and thereafter at regular intervals for a period of time specified by service agreements.
HAMILTON recommends that this verification take place four times yearly.
For volume verification in the field, HAMILTON will supply a verification tool, based on
gravimetrical measurements (the “gold standard” of volume verification). The detailed
specifications are listed in the specification table at the end of this chapter.
Conditions are valid only for the HAMILTON verification kit.
'LVSRVDO
After the life cycle of the instrument has terminated, the Microlab STAR may be shipped to
the original manufacturer or retailer. Otherwise local disposal regulations are to be observed.
7UDLQLQJ
Training in operation of the Microlab
STAR and the User Software will be provided by
HAMILTON personnel upon initial set-up.