5-9
METHODS
5.5.4 Predispensing Amount
The titration time can be shortened by adding a large fraction of the titrant at the start of the
analysis, if the approximate water content of the sample is known.
The predispensing amount can be set to deliver between 1 and 90% of the titrant required
to reach the end point.
Setting the amount to 0% will disable the titrant predispensing feature.
5.5.5 Pre-Analysis Stir Time
To avoid erroneous results or unreacheable endpoints when analyzing samples with limited
solubility, the sample must be completely dissolved in the solvent prior to the start of a
titration.
The pre-analysis stir time can be set between 0 and 1000 seconds. After the sample is added
to the reaction vessel the titrator will stir for the set period of time before any titrant (exclud-
ing predispensing) is added to the cell.