6-11
TITRATITRA
TITRATITRA
TITRA
TIONTION
TIONTION
TION
Averaging Sample Analysis Results
By pressing
, results will be added to the
Sample Analysis History
in order to obtain
an average of titration results.
Use the
and keys to scroll the concentration results list.
Use
to choose the sample concentration results that will be used for averaging.
Note: When there are no results selected dashes will appear in the Average Sample Concen-
tration and the Standard Deviation fields.
6.6 Titrant Standardization
While in Standby mode, press .
Note: If the drift value is zero, a warning message appears to inform the user that the
solvent may be overtitrated.