SETUP
Edit concentration point Measure Abs point
.
.
.
.
.
.
This procedure can be repeated until 10 calibration points have been added.
MANUAL ABS ENTRY
This allows users to enter the absorbance of standards with a known concentration. Up to 10 points can be used to calibrate the
method.
Press the CFM key to start the calibration.
Press the EDIT key to modify the concentration for the first standard.
Use the or key to highlight the digit to be modified. Press the or keys to set the desired value. Press
CFM to confirm the value or CLR to delete set value. Press the or key to abort calibration.
Press the CFM key to continue.