Safety requirem ents 
    EN 
Tablet1000 User manual                                                                                              Copyright © Qingdao Hantek Electronics Co., LTD 
4 
Installation (Overvoltage) Class This product is powered by a mains supply 
conforming to Installation (overvoltage) Class II. 
Warning: 
Ensure that no overvoltage (e.g. from lightning) reaches the product. Otherwise, 
the operator may be in danger of receiving electric shock. 
Installation (overvoltage) class definition 
Installation (overvoltage) Class I refers to the signal level applicable to the device 
measurement terminals connected to the source circuit where measures have been 
taken to limit the instantaneous voltage to the corresponding low level. 
Installed (overvoltage) Class II refers to the local distribution level for equipment 
connected to a mains (AC supply). 
The degree of pollution 
2 class 
Definition of pollution level 
⚫  Pollution degree 1: no pollution, or only dry non-conductive pollution. This 
pollution level has no impact. For example: clean room or air-conditioned office 
environment. 
⚫  Pollution level 2: generally only dry non-conductive pollution occurs. 
Transient conduction due to condensation may sometimes occur. For example: 
general indoor environment. 
⚫  Pollution level 3: conductive pollution occurs, or dry non-conductive 
pollution becomes conductive by condensation. For example, an outdoor 
environment with a canopy. 
⚫  Pollution level 4: Permanent conductive pollution through conductive dust, 
rain, or snow. For example: outdoor places. 
Security level 
Class 1 - Grounded products 
1.5  Maintenance and cleaning   
Maintenance: 
Do not expose the LCD to direct sunlight for a long time when storing or placing the 
oscilloscope. 
Cleaning: 
If the oscilloscope and probe are inspected frequently as required by operating